ESTRO 38 Abstract book
S922 ESTRO 38
EP-1712 Determination of photon output factors: implementation of the IAEA/AAPM TRS-483 Code of Practice P. De La Monja Rey 1 , C. Peraza 1 , M.L. Brosed 2 , A. Chaves 3 1 GenesisCare Madrid - Hospital San Francisco de Asís, Medical Physics Department, Madrid, Spain ; 2 GenesisCare Madrid, Medical Physics Department, Madrid, Spain ; 3 GenesisCare Spain, Medical Physics Department, Madrid, Spain Purpose or Objective The aim of this work is to determine the output factors (OFs) measured with six different detectors for an Elekta Versa HD linear accelerator equipped with the Agility head. The measured output factors were subsequently corrected by applying the factors proposed in the Code of Practice International Atomic Energy Agency 483 The OFs for field sizes ranging from 1.0 x 1.0 cm 2 to 10 x 10 cm 2 at a depth of 10 cm and at 100 cm source to detector distance were measured for 6 MV and 6 MV Flattening Filter Free (FFF). Several types of detectors, all from PTW-Freiburg, were used for comparison: microDiamond, diode-P shielded, diode-E unshielded, diode SRS - unshielded, PinPoint 3D ionization chamber and 0.125 cm 3 Semiflex ionization chamber. Off-axis dose profiles (in-plane and cross-plane) measurements were performed with the PTW BeamScan water phantom to ensure detector alignment and determine the effective field size. Output correction factors from IAEA/AAPM TRS- 483 were applied for all the detectors. In this study, the minimum field size analysed was 1.4 x 1.4 cm 2 for diode- p and 2 x 2 cm 2 for the chamber 0.125 cm 3 Semiflex. A daisy chaining approach was also utilised for the small OFs choosing the 4 x 4 cm 2 measurement as reference. Results For both energies, 6 MV and 6 MV FFF, the OFs measured for each detector as shown in Figure 1 and Figure 2. For the smaller field size of 1 x 1 cm 2 the maximum OF difference measured was 7.5% for 6 MV and 7.1% for 6 MV FFF, which, once corrected, lead to a maximum deviation respectively of 1.8% for 6 MV and 1.7% for 6 MV FFF. All detector response agree within 0.6 % except for the microDiamond where de OF measured were 1% higher than for other detectors. For field sizes greater than 2 x 2 cm 2 , the deviations of the corrected OFs were within 1% for both beams. The disparity between the direct and daisy– chain approach of OFs corrections was less than 0.3% for 6 MV and less than 0.5% for 6 MV FFF. (IAEA/AAPM TRS-483). Material and Methods
standalone device during patient treatments, the Discover can measure the MLC positions, gantry- and collimator- angle and compare these measurements to the planned values. Combined with the Delta 4 Phantom+ (DP+) for pre- treatment plan validation, the subsequent Discover-only measurements during patient treatment can then be used to calculate the dose distribution for each treatment on the DP+. A gamma analysis can be used to evaluate the delivered treatment fractions the same way the pre- treatment delivery was evaluated. Patient plans were developed according to our clinical protocol with two VMAT arcs. The attenuation of the Discover was determined for each photon beam energy during its commissioning and was accounted for in the patient plans. Each patient was localized and tracked using the Clarity US IG (Elekta). Tracking margins ensured that the PTV will be within 3 mm (maximum displacement) during beam-on time. We present five SBRT patients, with five fractions for each patient that were all delivered with the Discover in place. Additionally, the Discover and DP+ were used for all pre- treatment IMRT validations. We accumulated the five dose-fractions into the planning CT data set and compared the Discover measured results that were converted to the total dose outcome with the actual plan. Results To compensate for Discover attenuation, plan monitor units (MU) were increased by 1.3% for 6X flattened beams, 1.5% for 6X flattening filter free (FFF) beams and similarly for 10X for 1.0% and 10X FFF-beams for 1.2%. Figure 1 shows an example of the Discover gamma analysis with a criteria of 2%/2mm for the first fraction for one patient. As an example of our results, we found that the Discover device measured different gamma for each delivered fraction as shown in figure 2. The variation of gamma trended in line with the daily output difference.
Conclusion Using US IG tracking, we assure that our patients have less than 3 mm motion of the prostate during their whole treatment course. Similarly as the US IG tracking assures treatment delivery in the correct position, the Discover device assures the delivery of the radiation plan is as intended. The combination of these techniques gives much greater confidence to the clinician that the dose seen in the treatment plan is the dose that is being delivered to the patient.
Conclusion The results indicated that the diode-P overestimated OFs for fields smaller than 4 x 4 cm 2 . The microDiamond overestimated OFs for fields size smaller than 3 x 3 cm 2 ,
Made with FlippingBook - Online catalogs