2D vs 3D setup
2D vs 3D setup
33 patients, 100 paired CBCT-planar images Varian OBI
Phantom test: identical Detected error in CBCT is larger in all directions CBCT measurements better reproducible
2009
Fuller et al. 2009
Overall Conclusion 2D vs 3D setup correction Planar imaging underestimates the setup error, especially if you do it manually Deformation might contribute to this (Li et al., Fuller et al.)
Main reason for 3D setup
But is 3D enough?
How flexible is the neck ?
How flexible is the neck ?
• Purple: reference CT scan
• PTV conflicts with OAR
• You cannot align all at the same time: Compromise!
• Green: daily
localization Cone Beam CT scan
• Need for choices in alignment!
4