Gynae BT_Lucknow_2018

Calculation on MRI? It is OK!

 TG43 algorithm is based on water calculation and can be done on CT, MRI and US  Model based algorithms take tissue into account (based on CT), but has limited impact for gyn brachy

Implant

% Variation

9 ± 7 8 ± 8

Surface Mould (Nose)

Head and Neck (Base of Tongue)

8 ± 2.0 11 ± 14

Breast APBI –Multi Catheter

Lip Implant

Eye Lid 22 ± 37 Gynaecology – Vienna applicator (Polymer) 1 ± 0.2 Gynaecology – Ring applicator (SS) 4 ± 0.7

Courtesy Jamema Swamidas