IMRT 2018
courtesy P. Lang, C. Heinz (LMU Munich)
Precise dose calculation is still a challenge: Be aware of potential software BUGS!
Example:
depth dose curve in water for a central 1x1 cm field , 6MV results of a dose calc. algorithm (here: CC) of a widely used commercial TPS...
typical challenges for all dose calculation algorithms: small fields, irregular shaped fields (esp. off-axis), high energies (15MV and more)
measured depth dose curve
calculated depth dose curve, 1mm voxel size
calculated depth dose curve, 3mm voxel size
ESTRO IMRT course 2017 Dose Calculation in Static and Rotational IMRT Madrid April 13th, 2017
-6-