IMRT
Accuracy frame-based vs. frameless
1,50
1,00
0,50
Frame-based Frameless
0,00
Longitudinal Lateral
Vertical
-0,50 Average shift (mm)
-1,00
Overall 3D accuracy: 0.97 mm SD 0.24 mm (frame-based) 0.85 mm SD 0.52 mm (frameless)
37