Imaging for early stage breast cancer
IGRT with CBCT for prone APBI
343 CBCT verifications at 70 patients in prone position manual matching between planning and CBCTs using breast/chest wall interface
planning CT
CBCT
AP SI ML systematic 0.38 0.21 0.28 random 0.43 0.29 0.37 margin 1.26 0.73 0.96
the interfraction variations detected are comparable
to those reported for APBI with a supine setup
the currently applied margin to the CTV of 1.5 cm
adequately compensates for the detected setup
variability
Derived quantities from shifts in cm
Jozsef et al. IJROBP 81:568-574. 2011