In-room imaging and MR planning
Acoustic Noise
Manufacturer
Field Strength (T)
SPL (dB(A))
Lorentz force causes gradient coils to vibrate Sequence dependent Increases with B 0
Philips
1.5
112
Siemens
1.5
106
GE
1.5
110
Varian
3.0
118
Bruker
3.0
113
First
Painful at 134 dB Permanent damage at 120 dB
Field
thuMb
Motion
seCond
Current
F
= I dl B 0
L
Made with FlippingBook - professional solution for displaying marketing and sales documents online