In-room imaging and MR planning
Exposure Limits: RF
SAR
Whole body head Partial body* Local Body**
Local Extremity**
Normal
2 W/kg
3.2
2-10
10
20
Controlled
4
3.2
4-10
20
40
Research
> 4
> 3.2
> 10
> 20
> 40
Above are IEC 2010 limits with following notes: Over a time average 6 min; 10 s duration cannot exceed × 2 *scales as 10-8 × r where r is ratio of exposed mass to whole body mass **local SAR based on 10g tissue mass Assumes normal ambient temperature and humidity
Body Temperature
Max increase ( C)
Normal
0.5
Controlled
1.0
Research
> 1.0