In-room imaging and MR planning

Timing requirements

Q: how long does it take for a photon to travel 100 cm (typical bore size)?

A: c = 3.0×10 8 m/s → t = D/c = 3.3 ns

Thus, coincidence timing window cannot be reduced to less than 3-4 ns