Molecular Imaging 2017
PET hardware attenuation correction: First patients
HW + human µ- map
Comparison of FDG PET activity in n=10 HNC patients: • Reference scan • Scan with CH • Without AC • With AC
Pat 1: full AC
Deviation to reference scan (without CH):
Full AC of CH + coils
No AC of CH + coils
Tumor
0.6 % (-2.5 – 4.5)
-11.4 % (-14.0 – -6.2)
SUV mean
Brain
0.7 % (-0.8 – 2.3)
-14.0 % (-15.2 – -12.0)
Tumor
0.4 % (-3.6 – 5.8)
-11.8 % (-15.8 – -5.1)
SUV max
Brain
-1.4 % (-4.3 – 3.1)
-15.8 % (-19.0 – -10.3)
R. Winter, Tübingen