Physics Bucharest 2017

Interplay - Dynamic techniques

Problem : planning for IMRT, dynamic arc treatment or active scanning particle therapy on a moving target

→ the motion of the target may interfere constructively or destructively with the motion of the MLC leaves, the beam opening, the gantry rotation, the beam scan path and / or other dynamic parameters during treatment delivery. → The effect of such interferences is that the dose to sub-volumes in the patient may be significantly different from the planned dose, since interferences are not modeled in the treatment planning system.