Physics Bucharest 2017
Plan quality vs Delivery accuracy - VMAT
Leaf motion constraint
Gamma-analysis
• max distance a MLC leaf can travel between adjacent control points • ensures the deliverability of the optimized plan, but it also impacts the plan quality, delivery accuracy and efficiency • a less restrictive leaf motion constraint ( > 5 mm/deg) results in > plan quality but < accurate dose distribution • g -analysis passing rate from 98% to 80% when the allowable leaf motion increased from 3 to 20 mm/deg.
Chen F et al. Med Phys (38) 2011
Made with FlippingBook