Physics Bucharest 2017
10 MV X-rays and 10 MeV electrons
10 MeV electrons
10 MV photons
dose-scaling factor for electrons: 1.12
1. Irradiate film with electron and photons, use for electrons 1/1.12 * (prescribed dose) 2. Analyze film with sensitometric curve of photons Reliable relative dose distribution