Physics Bucharest 2017
Measured peripheral photon and neutron equivalent doses
Prostate IMRT plans
Peripheral equivalent dose: photons % of tumor dose
Peripheral equivalent dose: neutrons % of tumor dose
1.5
1.5
1.0
1.0
6MV 10MV 15MV 18MV
6MV 10MV 15MV 18MV
0.5
0.5
0.0
0.0
Equivalent dose
0
20
40
60
0
20
40
60
distance from central axis [cm]
distance from central axis [cm]
Neutrons: important for energies > 10 MV For 18 MV, neutrons are more important than photons for large distances
Kry et al 2005
14
Made with FlippingBook