Physical Parameter
Suspension Level
Reference
Type
(IEC (2007, 2008c) clause numbers
unless stated)
Symmetry of electron fields (max/min ratio) Maximum ratio of absorbed dose (max/min ratio)
>1.05
A
1.09 See IEC
A
Dose monitoring system
7
Weekly calibration check
>2 %
A
>0.5 %
A
Reproducibility
Proportionality
>2 %
A
Dependence on angular position of gantry and beam limiting device
>3 %
A
Dependence on gantry rotation
>2 % - electron radiation >3 % - X-radiation
A
Stability throughout the day
>2 %
A
Stability in moving beam radiotherapy
See IEC
A
8
Depth dose characteristics
X-radiation
Criteria for Acceptability of Medical Radiological Equipment used in Diagnostic Radiology, Nuclear Medicine
Penetrative quality
>3 % or 3 mm.
A
and Radiotherapy. RP 162 (European Commission), 2013
>2 %
IPEM (1999) B
Depth dose and profiles
Made with FlippingBook - Online Brochure Maker