Protective shielding Collimator channels Frame-based Stereotactic Radiosurgery Positioning Accuracy Accuracy and stability of positioning in radiosurgery: long –term results of the Gamma Knife system. Heck B et al
Leksell ® Coordinate Frame Isocenter/ Target in the brain
Graf Chromic films densitometric measurements X: - 0.014+/- 0.09mm Y: 0.013+/- 0.09mm Z: - 0.002+/- 0.06mm MRI-based target definition
Patient positioning system
Radiation sources
X: 0.06+/-0.09mm Y: 0.04+/-0.09mm
All measured data were within a sphere of 0.2mm radius Target delineation: GTV=PTV
Med Phys 2007 Apr; 34(4): 1487-95