SBRT 2016
New Skills and New Knowledge: Training
+1.6
+0.8
Calibrate positions of all nodes At each node, the isocrystal is scanned in a grid pattern perpendicular to the incident beam Light intensity map is recorded Two 5 x 5 grids (coarse and fine) The calibrated node corresponds to the center of the light intensity map Must be performed separately for fixed cones and IRIS Result: Node “error list”1
0
-0.8
-0.8
+0.8
nominal node 1 st iteration (coarse) calibrated node
-1.6
-1.6
[mm]
+1.6
Adapted from Christoph Fürwerger
Made with FlippingBook