Wrong CT for matching: 4D-CT

• 4D-CT used to generate Midvent plan. • CTV delineation at Midvent position used to generate PTV and position isocenter • (Plan calculated on average CT) • CBCT should be matched on midvent CT • However, average CT was used, introducing systematic shift! (planned CTV position <> CTV position on reference CT)