SBRT2015
Pre-SABR programme
• Consider immobilisation for each SABR site • Audit set-up uncertainties for system used for non-SABR patients • Calculate systematic and random uncertainties • Assess if appropriate for SABR by comparing with suggested SABR margin • Adapt as necessary e.g. chin strap, vac bag
Population (cm) vert syst vert rnd long syst long rnd lat syst
lat rnd 0.094044 0.14978 0.202467 0.406479 0.152902 0.269381
Standard lung immobilisation 35 patients
Margins (mm)
Vert
Long
Lat 5.7
3.4
7.9
Standard lung = 9mm PTV margin
SABR lung = 5mm PTV margin
September 2015
ESTRO SABR course
5