SBRT2015
IGRT Considerations (2)
• RTT led process : advanced IGRT competency • Training and competency assessment
Example from RMH: A 4 stage training programme was implemented including: • Training session for RTT on good practice for SABR verification • Competency based testing of matching 20 offline CBCT lung SABR verifications. • To be assessed as competent a 90% concordance with clinicians verification was required. An acceptable match was translocation < 2mm and rotation < 2 ° compared to the clinicians. To audit, a clinician retrospectively verified 20 patients lung SABR CBCT images matched by RTT .
Audit of Advanced Competency for Radiographer led lung Stereotactic Ablative Body Radiotherapy (SABR) verification. David Frost et al. Royal Marsden NHS Foundation Trust.
September 2015
ESTRO SABR course
14
Made with FlippingBook