SBRT2015
Towards extracranial SRS: body frames
… still requires IGRT
Deviations of 12 mm have been observed Applying a safety margin of 5 mm, 12-16% of the target might be partially missed.
(Wulf et al. )
Stereotactic Body Frame, Lax et al.
• AAPM TG 101 recommendation: ! “Body frames and fiducial systems are OK for immobilization and coarse localization” ! “They shall NOT be used as sole localization technique”
12
SBRT 2015 - D. Verellen