ESTRO 2024 - Abstract Book
S3249
Physics - Detectors, dose measurement and phantoms
ESTRO 2024
Phases
Static (cGy)
Moving (cGy)
Difference (%)
1
248.95
248.20
0.30
2
247.95
247.21
0.30
3
287.01
286.27
0.26
4
285.01
285.27
-0.09
6
284.02
285.01
-0.35
8
285.77
285.27
0.17
The radiochromic films have been compared with each other with a gamma analysis of 3% 1mm, 2% 1mm and 1% 1mm with a threshold of 10%, yielding a gamma of 3% 1mm >99% for 3 phases and a gamma of 3% 1mm >95% for 6 phases (figure 2).
Figure 2: Gamma analysis with EBT3 radiochromic films of 1, 3 and 6 phases of the respiratory cycle.
As shown in figure 2, the gamma analysis of the different phases of the respiratory cycle shows a result of 3% 1 mm > 95%, however the best results were obtained for the 3-phase treatment.
For few phases tratments, the ITV generated is smaller, helping to better preserve organs at risk. However, this requires a longer treatment time in the accelerator, which could be a drawback in the actual treatment of patients.
Conclusion:
In this work it has been verified that the irradiation of a treatment with gating in the accelerator versus a continuous treatment corresponding to a treatment without respiratory control introduces a difference in the delivered dose due