Physics - Detectors, dose measurement and phantoms
ESTRO 2024
to treatment interruption and resumption of less than 1%. The differences in the dose distribution coincide in at least 95% of the points within a gamma of less than 3% and 1 mm, even for a treatment with 6 phases.
Acknowledgements
This Project has received funding from Euratom research and training programme 2019-2020 under agreement No. 945196 (SINFONIA Project).
Keywords: Gating, Radiochromic Film, Phantom
References:
1. Lee, S. H., Kim, K. B., Kim, M., Yoo, H. J., Park, S. W., Jung, H., Ji, Y. H., & Yi, C. (2013). Quantitative analysis of dose distribution to determine optimal width of respiratory gating window using GAFChromic EBT2 film. Journal of the Korean Physical Society, 62(4), 657-663.
2.Savanović, M., Jaroš, D., &Foulquier, J. N. (2021). Comparison of phase-gated and amplitude-gated dose delivery to a moving target using GAFChromic EBT3 film. Journalof Medical Physics.
3.Crijns, S., Kok, J. G. M., Lagendijk, J. J. W., &Raaymakers, B. W. (2011). Towards MRI-guided linear accelerator control: gating on an MRI accelerator. Physics in Medicine and Biology, 56(15), 4815-4825.
1246
Digital Poster
Detection of submillimeter MLC leaf positioning error through log files analysis using Mobius3D
Nina Cavalli 1 , Elisa Bonanno 1 , Giuseppina Borzì 1 , Alessia D'Anna 2 , Martina Pace 1 , Giuseppe Stella 2 , Lucia Zirone 1 , Carmelo Marino 1 1 Humanitas, Medical Physics, Catania, Italy. 2 University of Catania, Department of Physics and Astronomy E. Majorana, Catania, Italy
Purpose/Objective:
With the purpose to establish the strength of our calculation based pre-treatment patient Quality Assurance (QA) method for HyperArc (HA) treatments, we checked the ability of Mobius 3D system to detect intentional errors ≤ 1mm related to the high definition MLC position through the log files analysis in MobiusFX (MFX) Module.
Material/Methods: