S3706
Physics - Dose prediction, optimisation and applications of photon and electron planning
ESTRO 2024
plan A and it was 86.7% to 100% (mean = 95.5%, Std Dev = 4.71%) for plan B. In S-I direction, which lung tumor moves the mostly, the coverage remains the same or even improved if iso moves superiorly for both plans, scenario 4. The tumor coverage decreased to 89.4% from 99.5% if iso moves inferiorly for plan A and it decreased to 99.7% from 100% for plan B, scenario 3. Fig. 2. is a sagittal view of the representative iso dose distribution of the original plan A and B and plans of scenario 3 for plan A and plan B.
Isocenter shift (cm)
V95 to tumor (%)
Scenario R-L
I-S
P-A
Plan A
Plan B
0
0.0
0.0
0.0
99.5
100.0
1
-0.5
0.0
0.0
96.9
97.0
2
0.5
0.0
0.0
85.2
86.7
3
0.0
-0.5
0.0
89.4
99.7
4
0.0
0.5
0.0
100.0
100.0
5
0.0
0.0
-0.5
93.2
92.5
6
0.0
0.0
0.5
90.7
92.6
Fig. 2. The iso dose plans of 2 selected senecios of plan A and B. top left: plan A, zero shift. top right: plan A 5 mm shift inferior. bottom left: plan B, zero shift; bottom right plan B 5 mm shift inferior.