S.H. Jeon 1 , S.Y. Park 1 , J.H. Kim 1 , J.I. Kim 1 , J.M. Park 1 1 Seoul National University Hospital, Radiation Oncology, Seoul, Korea Republic of Purpose or Objective To suggest an optimal planning target volume (PTV) margin in stereotactic body radiotherapy (SBRT) of the spine. Material and Methods From December 2014 to July 2016, 40 patients received 42 fractions of SBRT for spinal tumors to thoracic or lumbosacral spines using a volumetric modulated arc therapy technique and patient immobilization. Before treatment, kilovoltage cone-beam CT (CBCT) images were obtained for a 4 degrees of freedom (DoF) correction of patients alignment (translation + yaw). After corrections were made, additional CBCT was acquired just before treatment delivery (pretreatment CBCT). Immediately following SBRT, CBCT was acquired again (posttreatment CBCT). Residual setup errors for pretreatment CBCT was determined by a 6 DoF manual matching. Intrafraction motions were calculated as differences in errors between pretreatment and posttreatment CBCT. Three clinical target volumes (CTVs) were generated by translating and rotating original CTV by residual setup errors alone (CTV_R), intrafraction motions alone (CTV_I), and residual setup errors and intrafraction motions combined (CTV_R+I), respectively. Adding various uniform margins to original CTV generated PTVs. The impact of PTV margins on CTV coverage was evaluated. A provisional criterion of adequate CTV coverage was that PTV encompasses at least 97% of CTV. Results Time interval between pre-treatment and post-treatment CBCTs was 6.8±2.5 min (mean±2SD). The 2SD values of lateral, vertial, longitudinal translations and pitch, roll, and yaw were 0.7mm, 0.8mm, 1.1mm, 1.7°, 1.1°, and 1.6°for residual setup errors and 1.0mm, 0.9mm, 0.9mm, 1.1°, 0.8°, and 1.1°for intrafraction motions, respectively. Without margins, PTV showed adequate coverage for CTV_R, CTV_I, and CTV_R+I in 48% (20/42), 71% (30/42), and 48% (20/42) of fractions, respectively. With 1-mm uniform margins, PTV was adequate for 95% (40/42), 98% (41/42), and 100% (42/42) of fractions, respectively. 2-mm uniform margin was adequate in all fractions for all three CTVs. Conclusion With appropriate immobilizations and 4DoF corrections, a uniform 1-mm PTV margin may ensure an adequate CTV coverage in most treatment sessions of spine SBRT. Combined with a shortened treatment time, the small extent of intrafraction motions may obviate the need of treatment interruption for additional intra-session image guidance. Despite perfect 6 DoF patient alignment, 1-mm PTV margin is still needed to address intrafraction motions. PO-0864 Accuracy of fiducial based correction of target motion in prostate SBRT treatments T. Viren 1 , M. Korhonen 2 , J. Seppälä 1 1 Kuopio University Hospital, Cancer Center, Kuopio, Finland 2 University of Eastern Finland, Department of Applied Physics, Kuopio, Finland Purpose or Objective Robotic stereotactic body radiotherapy (SBRT) incorporating a fiducial based motion tracing system has enabled almost real-time correction of intra-fraction motion of a prostate during SBRT treatments of prostate cancer. However, the effect of number and positioning of the fiducials and the amount of prostate movements on the accuracy of the treatment has not been reported. The aim of the present study was to investigate the accuracy of the fiducial based correction of target motion in
(NKI).
PO-0862 Correlation of Liver and Pancreas Tumor motion with Normal Anatomical Stru ctures R. Kaderka 1 , A. Paravati 1 , R. Sar kar 1 , J. Tran 1 , K. Fero 1 , N. Panjwani 1 , D. Simpson 1 , J. Murphy 1 , T. Atwood 1 1 University of California San Diego, Department of Radiation Medicine and Applied Sciences, San Diego, USA Purpose or Objective Target motion caused by respiration remains the central challenge to delivering SBRT in the abdomen. For targets in the pancreas and liver, SBRT oftentimes necessitates placement of metal fiducials to determine tumor position with fluoroscopy, due to difficulty in visualizing tumors on non-contrast imaging. Metal fiducials have limitations in that they represent an invasive procedure which can introduce treatment delays. Furthermore, fiducials can migrate from their intended position, and the metal can introduce imaging artifacts which make tumor delineation a challenge. We hypothesized that upper abdominal tumor motion would correlate with the motion of nearby organs and could thereby serve as a fiducial-less proxy for tumor motion. Material and Methods Fifteen patients (12 with pancreas and 3 with liver tumors) underwent a 4-dimensional (4D) CT simulation prior to treatment with SBRT. 4D CT images were divided into 10 phases and normal tissues were contoured on a single 4D- CT phase and propagated to the other phase s using deformable image registration. As a means of quality control for image registration and contour propagation the liver was manually contoured on all phases for 5 patients by physicians and compared to the automated contour propagation using a Dice coefficient. Motion was defined from the center-of-mass of each structure, and a patient- specific linear tumor position prediction model based on liver position was developed. Results We found a strong overlap of manually entered contours and the automatically segmented contours with a mean Dice-coefficient of 0.95 (standard deviation 0.01). The linear models accurately predicted tumor motion with a mean absolute error of 0.5 mm and no error greater than 3.0 mm. Mean absolute and maximum errors by direction and tumor type are listed in the table below.
Anterior- posterior direction
Superior- inferior direction
Left-right direction
Pancreas tumors mean absolute error (mm) Pancreas tumors maximum error (mm) Liver tumors mean absolute error (mm)
0.3
0.4
0.5
1.0
1.7
2.6
0.3
0.4
0.8
Liver tumors maximum error (mm)
1.4
2.7
3.0
Conclusion This study demonstrates that normal organ motion could serve as a fiducial-less proxy for tumor motion with SBRT in the upper abdomen when on-site real-time 4D volumetric imaging becomes available during treatment. Deformable image registration has been demonstrated to be a reliable and fast tool for segmentation of normal organs. Moving this motion management approach into clinic requires additional research to optimize 4D image quality and understand inter-fraction reproducibility. PO-0863 Suggestion of optimal planning target volume margins for stereotactic body radiotherapy of the spine