S464 ESTRO 36 2017 _______________________________________________________________________________________________

this study was to develop and validate a 4D-MRI guided mid-position (midP) correction strategy on an MR-Linac. Material and Methods Experiments were performed on an MR-Linac (ATL1, Elekta AB, Sweden), using the CIRS MRI-LINAC Dynamic Phantom (CIRS Inc., USA). The moving cylinder was filled with anisotropic MRI contrasts and a Perspex spherical target. Motion was performed in CC direction using a Lujan 4 motion pattern with a 20mm amplitude and 4s period. First, a T2-weighted MRI scan was acquired in midP. The cylinder and target were segmented and the target was expanded with a non-uniform margin (LR, AP:10mm; CC:20mm). A density overwrite of 1 was assigned to the structures and a treatment plan consisting of a single anterior beam shaped around the PTV was created in Monaco (Version 5.19.01 Research). Then, baseline shifts in CC direction of 5, 10, 15 and 20mm were applied to the phantom motion. For every shift, a retrospective self- sorted 4D-MRI was acquired (axial single-shot TSE, 2x2x5mm 3 , TE/TR=60/400ms, 30dyn) and each phase was registered to the midP reference image to calculate the time average displacement. The plan was adapted accordingly, performing a virtual couch shift (simple dose shift) using aperture morphing in Monaco. All plans were delivered while electronic portal imaging device (EPID) cine images were acquired. The time average displacement of the target was calculated from the EPID images and geometric accuracy of the workflow was quantified as the distance of the average position of the target to the field edges in the EPID images. Results In Figure1, MRI and EPID images of the midP and a shifted inhale and exhale position are shown. Table1 shows the time average displacement of the target in the 4D-MRI and the EPID images with respect to the reference as well as the distance of the average target position to the field edges. The geometric accuracy of the 4D-MRI guided workflow was 0.3±0.4mm in CC, which includes the 4D-MRI registration accuracy.

PO-0861 Geometric validation of a 4D-MRI guided correction strategy on the MR-Linac T. Van de Lindt 1 , R. Koopman 1 , A. Van de Schoot 1 , I. Torres-Xirau 1 , U. Van der Heide 1 , J.J. Sonke 1 1 Netherlands Cancer Institute Antoni van Leeuwenhoek Hospital, Radiation Oncology, Amsterdam, The Netherlands Purpose or Objective Currently in radiotherapy, respiratory motion correction strategies are performed by the use of 4D-(CB)CT. However, moving targets in for example the upper abdomen are not (clearly) visible on these images because of low soft-tissue contrast. The introduction of an integrated MRI and linear accelerator (MR-Linac) will allow for daily MRI-guidance of the tumor. Therefore, the aim of

Conclusion 4D-MRI guidance on an MR-Linac was shown to be feasible and had sub-millimeter accuracy. Such a correction strategy has great potential for moving targets that are difficult to visualize on alternative image guidance modalities. Acknowledgements: This research was partly sponsored by Elekta AB, Stockholm, Sweden. The authors would like to thank CIRS Inc., Robert Spaninks (Elekta) and Jochem Kaas, Natasja Janssen, Ben Floot and Marco van den Berg

Made with