ESTRO meets Asia 2024 - Abstract Book
S391
RTT – Treatment planning, OAR and target definitions
ESTRO meets Asia 2024
We selected 4 patients with hip prostheses (3 unilateral and 1 bilateral) and created 4 plans with different planning strategies:
- Plan_1: No restriction.
- Plan_2: "No entry+exit" optimization.
- Plan_3: "No entry" optimization.
- Plan_4: Exclusion segments.
Figure 1. Patients
Results:
We observed a considerable decrease in PTV coverage in Plan_2 for all patients due to its highly restrictive condition. However, in Plan_3, we achieved good PTV coverage and dose conformity without affecting the organs at risk.
All patients received 2 full arcs except for Patient 1, who received two partial arcs. Therefore, Plan_4 was discarded for this patient since there were not enough arcs to obtain optimal results.
Plan_4 was tested for Patients 2 and 3 using segments of 35º-90º and 63º-138.7º in both clockwise (CW) and counterclockwise (CCW) directions. Meanwhile, for Patient 4, the segments were 247º-305º CW and 99º-31º CCW.
Table 1. Measures
In cases where angles coincide with both the prosthesis and partial PTV, it is better to use Plan_3 because it continues to irradiate other areas of the PTV without entering the prosthesis, unlike Plan_4.
Conclusion: