Germany 6 National Center for Tumor Diseases NCT, partner site Dresden, Dresden, Germany Purpose or Objective Dual-energy CT (DECT) provides additional patient information to potentially improve delineation and range accuracy in proton therapy. Motion during sequentially acquired DECT scans might hamper its reliability. Here, we analysed the clinical feasibility of sequential 4D DECT scans (2 consecutive respiratory correlated 4D CT scans) for non-small cell lung cancer (NSCLC) patients and its applicability for proton dose calculation. Material and Methods For 3 advanced stage NSCLC patients with maximal tumor motion of 1mm in cranio-caudal direction, 4D DECT scans were sequentially acquired during the course of treatment with a Siemens single-source DECT scanner. 80/140kVp average CT datasets and 4 breathing phases (relative amplitude sorting) were reconstructed and compared visually. These DECT datasets were further processed in syngo.via (Siemens Healthineers) to calculate 79keV pseudo-monoenergetic CT (MonoCTs) and stopping-power-ratio datasets (SPR, derived from electron density and photon cross section), Fig.1a. Passively scattered proton treatment plans were recalculated on MonoCT and 140kVp datasets using the clinical heuristic CT-number-to-SPR conversion (HLUT). Furthermore, worst-case scenarios using a single proton beam covering artificial target volumes encompassing the diaphragm (13-23mm motion) were generated. Dose distributions derived from MonoCT and 140kVp datasets were compared with 2D gamma analyses (0.1% dose and 1mm geometrical difference) to validate DECT image post processing. Finally, a patient-specific DECT-based SPR prediction was applied on 4D DECT datasets and followed by dose calculation to assess proton range shifts compared to the MonoCT-based HLUT approach. Results Visually, no differences between the two sequential 4D DECT scans were found. Breathing patterns did not change more between the 2 scans than within a single scan. Clinical dose distributions on MonoCT and 140kVp datasets were similar with an average gamma passing rate of 99.9% (99.2%-100%). The maximal dose difference was 0.8%, Fig.1b. The worst-case scenario plans had a minimal passing rate of 92.4% (average 99.3%) with maximal dose difference of 3.3%. Using the MonoCT dataset with clinical HLUT instead of the DECT-based SPR dataset for dose calculation led to clinically relevant mean range shifts (±1SD) of 2.3(±0.8)%, Fig.2.

Material and Methods As a first step, the effect of the CT contrast medium Imeron® 300 (Bracco Imaging Deutschland GmbH, Germany) on RED/SPR determination was investigated in a dilution series over a range of iodine concentrations between 0.3 and 300 mg/ml. CT images were acquired on a Somatom Definition Flash dual-source CT scanner (Siemens Healthineers, Forchheim, Germany) in single- energy (SECT, 120 kVp) and dual-energy (DECT, 80/140Sn kVp) scan mode. RED and SPR images were obtained (a) from SECT datasets by applying the respective calibrated Hounsfield look-up table and (b) from DECT datasets using the software application syngo.CT Rho/Z (Siemens) and an SPR calculation scheme previously validated by the authors in phantoms, biological material and patients. Results Calculating RED/SPR from a DECT dataset with typical contrast enhancement (max. 160 HU at 120 kVp corresponding to 6 mg iodine per milliliter) could limit the impact on both RED and SPR to 1% compared to 5-10% when using a contrast-enhanced SECT image (Figure 2). Consequently, dose calculation could be performed directly on DECT-derived RED/SPR images.

Conclusion Dose calculation on a RED/SPR dataset derived from a contrast-enhanced dual-source DECT scan is conceivable. This can make additional native scans obsolete, thereby simplifying the treatment planning workflow and lowering the patient dose by 50% (one instead of two scans). A clinical trial is currently underway to investigate the role of contrast-enhanced DECT for patient radiotherapy planning. EP-2090 Clinical feasibility of 4D single-source dual- energy CT for proton therapy of lung-cancer patients P. Wohlfahrt 1,2 , C. Hofmann 3 , E.G.C. Troost 1,2,4,5,6 , C. Richter 1,2,4,5 , A. Jakobi 1,2,4 1 OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus- Technische Universität Dresden- Helmholtz-Zentrum Dresden - Rossendorf, Dresden, Germany 2 Helmholtz-Zentrum Dresden - Rossendorf, Institute of Radiooncology - OncoRay, Dresden, Germany 3 Siemens Healthineers, Radiation Oncology, Forchheim, Germany 4 Department of Radiotherapy and Radiation Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus- Technische Universität Dresden, Dresden, Germany 5 German Cancer Consortium DKTK, partner site Dresden- and German Cancer Research Center DKFZ, Heidelberg,

Made with FlippingBook - Online magazine maker